Introduction
A
complete
procedure
operational
standards and calibration of the
Type L Plug-In Unit is provided in this section
o f
the manual.
The steps in the procedure
are
arranged in correct sequence to avoid
unnecessary repetition.
The step-by-step instructions also furnish
an orderly approach for isolation of malfunctions
which may develop.
cedure
should be used in conjunction with
any maintenance and troubleshooting system.
If
an abnormal
indication
some point in the procedure, it is not usually
necessary to locate its cause before continuing
to the next step. Additional symptoms revealed
by perform ing further steps w ill frequently
sim plify the task of locating trouble.
F or this procedure, the L Unit should be
used in conjunction with a properly calibrated
Tektronix Type 540- o r 550-Series O scillo
scope. Control settings are listed by the names
usually
applied to them on the scopes with
which the Type L Unit may be used. Alternate
names are in parentheses.
Test
equipment used in a particular step
should be left connected at the end of that
step
unless
instructions
Sim ilarly, controls not mentioned are assumed
to
be
in the positions they were in at the
conclusion of the preceding step.
If information explaining the normal operation
o f front-panel controls is needed before starting
the procedure, consult the Operating Instructions
section of this manual.
AA
fo r
checking the
Consequently, this pro
is
obtained at
state
otherwise.
Calibration Procedure - Type L Plug-In
C A L I B R A T I O N
P R O C E D U R E
Equipment Required
The following equipment or its equivalent
is required to perform a complete calibration
o f the Type L Plug-In Unit.
1. Test Oscilloscope, Tektronix 540- or 550-
Series.
2. Square-wave generator, Tektronix Type 105.
Required specifications are: 13-nsec risetim e
o r less; output frequencies of approximately
1 kc
and 10 kc; output amplitude variable
from 10 to 100 volts across 600-ohm internal
load.
3. Square-wave generator, Tektronix Type 107.
Required specifications are: 3 nsec risetim e
o r less; output frequency of approximately 1 MC;
output amplitude variable to at least 0.35 volt.
4. Sine-wave generator, Tektronix Type 190( )
Constant-Amplitude Signal Generator. Required
specifications are: frequency range of 50 kc
to
13 MC;
output amplitude of 0.14 volt or
m ore;
output must be
o r automatically) fo r constant amplitude within
the above frequency range. Type 190( ) denotes
any model of the Generator.
5. Calibrated volt-ohmmeter (VOM), with sen
sitivity of at least 20,0000/volt.
6 . 2 - 500 coaxial cable, Tektronix Type P52
7 . 1 - 500 Terminating Resistor
8. 1 - 20 pf
Input Capacitance
9. 2 - 500 X10T Attenuator
10. 1 - 500 X5T Attenuator
O I N J
5
adjustable (manually
012
001
(
-
)
(011-045)
Standardizer
011
022
(
-
)
(011-031)
(011-032)
5-1