Teledyne 3000TB Operating Instructions Manual page 74

Trace oxygen analyzer
Hide thumbs Also See for 3000TB:
Table of Contents

Advertisement

Trace Oxygen Analyzer
interval (6 months is the default setting) after which a notification on the
Analyze screen will advise that sensor maintenance is needed. This
function can be set so that in addition to the notification, it also triggers
the system failure alarm. It can also be turned off.
To access the Maintenance function use the Use the ◄ / ► arrow
keys to navigate to the fourth System menu.
Use the RIGHT button to move the blinking to Maintenance and
then press Enter to go to the next screen.
The timer in the maintenance function is set in months (M) and
days (D) on the top line of the display. The maximum setting is M=30
and D=30. The default is M=6 and D=0. The user has the option of
turning the function On or Off or to have it trigger the system alarm
(Alarm).
To change the notification period, use the RIGHT arrow key to
move the blinking to the appropriate field M(onth) or D(ay) and then use
the ▲/▼ keys to increment or decrement the value. Press Enter when
the appropriate setting is displayed.
The second line is the actual countdown and displays the time left
before a notification or alarm will occur. It counts down until the
Maintenance warning is set. To re-start the counter the operator must
select "Reset" and press Enter. The operator is prompted to press Enter
again to re-affirm the restart of the counter.
Pressing ESC will return you to the previous screen.
Teledyne Analytical Instruments
Maintenance Serial#
Stream SerMode More
Set: M: 0 D: 7 On
Nxt: M: 0 D: 6 Reset
Reset Maint Timeout?
<ENT> Yes <ESC> No
Operation
59