Fixed Range Analysis - Teledyne 3000TB Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3000TB:
Table of Contents

Advertisement

Trace Oxygen Analyzer
Use the ◄ / ►arrow keys to blink the range to be set: low (L),
medium (M), or high (H).
Use the ▲/▼ arrow keys to enter the upper value of the range (all
ranges begin at 0 ppm). Repeat for each range you want to set. Press Enter
to accept the values and return to Analyze mode. (See note below.)
Note: The ranges must be increasing from low to high, for
example, if range 1 is set as 0–100 ppm and range 2 is set
as 0–1,000 ppm, range 3 cannot be set as 0–500 ppm
since it is lower than range 2.
Ranges, alarms, and spans are always set in ppm units (over the
entire 0-250,000 ppm range), even though all concentration-data outputs
change from ppm units to percent when the concentration is above
10,000 ppm.

4.6.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing
the analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, enter the range
function by pressing the Range button on the front panel.
Use the ◄ / ►arrow keys to move the blinking over AUTO.
Use the ▲/▼ arrow keys to switch from AUTO to FX/LO, FX/M ED,
or FX/H I to set the instrument on the desired fixed range (low, medium,
or high).
Press Escape to re-enter the Analyze mode using the fixed range.
Teledyne Analytical Instruments
L—100
M—1000
H—1 0000 Mode—FX/ LO
or
L—100
M—1000
H—1 0000 Mode—FX/MED
or
L—100
M—1000
H—1 0000 Mode—FX/ HI
Operation
57