15 EMC Information

portable and mobile RF communications equipment and the ME equipment
TheAlternating Pressure Air Flotation Mattress is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of theAlternating Pressure Air Flotation Mattress can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
Issue 9– 17122020
Recommended separation distances between
Alternating Pressure Air Flotation Mattress
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
 
3,5
 
P
d=
 
  
1
0.12
0.37
1.17
3.7
11.67
m
80 MHz to 800 MHz
 
3,5

P
d=
 
  
1
0.12
0.37
1.17
3.7
11.67
as
800 MHz to 2.5 GHz


7

P
d=
 
  
1
0.23
0.74
2.33
7.37
23.33
Page 16 of 20