Audio Volume; Activation Tone; Alarm Tone; Return Electrode Sensing - Bovie SPECIALIST PRO User Manual

Electrosurgical generator
Hide thumbs Also See for SPECIALIST PRO:
Table of Contents

Advertisement

Audio Volume

The audio levels stated below are for activation tones (bipolar, cut and coag) and alarm tones (return electrode and
system alarms) at a distance of one meter. Alarm tones meet the requirements for IEC 60601-2-2.

Activation Tone

Volume (adjustable)
Frequency
Duration

Alarm Tone

Volume (not adjustable)
Frequency

Return Electrode Sensing

The system presents audible and visible alarms when it senses no return electrode.
Single Plate
Split Plate

Low Frequency (50-60 Hz) Leakage Current

Enclosure source current, ground open
Source current, patient leads, all outputs
Sink current at high line, all inputs
User's Guide • Specialist | PRO
40 to > 65 dBA
Cut: 610 Hz ± 10 Hz
Blend: 610 Hz ± 10 Hz
Pinpoint: 840 Hz ± 10 Hz
Spray: 840 Hz ± 10 Hz
Bipolar: 840 Hz ± 10 Hz
Continuous while the generator is activated
65 dBA at a distance of one meter
2.44 kHz / 450 ms / 1.22 kHz / 450 ms
Trip resistance: 0
Continuous measurement:
Once the system establishes the single-plate electrode resistance, an
increase of 20
alarm condition exists, the system deactivates output power.
Trip resistance: 10
Continuous measurement:
Once the system establishes the split-plate electrode resistance, an
increase of (35 ± 5)% in resistance will cause an alarm. When the alarm
condition exists, the system deactivates output power.
< 500 µA 220 - 240 VAC
< 300 µA 90 - 120 VAC
Normal polarity, intact ground: < 10 µA
Normal polarity, ground open: < 50 µA
Reverse polarity, ground open: < 50 µA
< 50 µA
to 8
± 1
± 25
in resistance will cause an alarm. When the
± 5
to 135
± 10
43