Warnings - Kodak X-Omat 5000 RA PROCESSOR Diagnostics

Hide thumbs Also See for X-Omat 5000 RA PROCESSOR:
Table of Contents

Advertisement

Warnings

Error
Error
Code
Description
E128
Top cover not in
place
E129
Tanks currently being
filled
E130
Replenish pumps
disabled
E132
Developer under set
temperature
E133
Developer over set
temperature
E134
Dryer under set
temperature
E137
Film accumulator
LED error
E141
Low developer tank
level
E142
Low fixer tank level
5B6342 – January 1996
Possible
Malfunction
When this error occurs, the following functions are disabled:
• film transport
• film accumulation
The TOP COVER is opened. Close the TOP COVER.
INTERLOCK SWITCHES S3
and S4
• MAGNETS
7000 BOARD
When this error occurs,
• the following functions are disabled:
– film transport
– film accumulation
• the following parts are disabled:
– RECIRCULATION PUMP
– 3 HEATERS
None
None
None
None
None
6000 BOARD
When this error occurs, the:
• RECIRCULATION PUMP is disabled
• temperature control for the fixer and developer is disabled
None
When this error occurs, the:
• RECIRCULATION PUMP is disabled
• temperature control for the fixer and developer is disabled
None
Action
Check S3 and S4. If necessary, install a new
SWITCH or MAGNET.
Install a new BOARD.
None. This message will clear automatically.
Use the KEYPAD to select either Automatic or
Flooded Replenishment to enable the PUMPS.
None. This message will clear automatically when
the developer reaches the setpoint temperature.
None. This message will clear automatically when
the developer reaches the setpoint temperature.
None. This message will clear automatically when
the DRYER reaches the setpoint temperature.
Check that no chemical deposits are on the
PROTECTIVE COVER for the BOARD.
De-energize and energize the PROCESSOR to
reset the:
• MICROPROCESSOR on the 5000 BOARD
• MICROPROCESSOR on the 6000 BOARD
Check the BOARD. Use the internal diagnostics. If
necessary, install a new BOARD.
None. This error will clear automatically when the
developer solution reaches the correct level.
None. This error will clear automatically when the
fixer solution reaches the correct level.