Fora VSM100 Pro Operator's Manual page 6

Vital signs monitor (blood glucose, blood pressure, temperature, spo2 and ecg measurement)
Table of Contents

Advertisement

CAUTION!
When several blood pressure measurements are taken on the same patient, regularly
check the cuff site and extremity for possible ischemia, purpura, and/or neuropathy.
Wrapping the cuff too loosely may lead to inaccurate results.
ECG Warnings and Cautions
These warnings and cautions pertain to the ECG warnings and cautions.
WARNING!
In order to minimize interference and the danger of burns to the patient, only use
FORA ECG cables. Keep the ECG cable as far away as possible from any
electrosurgical cables. Make sure that the electrosurgical return conductor (neutral) is
properly attached to the patient.
Danger of destroying the monitor during defibrillation! The monitor is type CF
protected only when the original FORA patient cables are used.
Patients with a pacemaker must be observed continuously because the heart rate
from the pacemaker might still be registered in case of a cardiac arrest or some
arrhythmias. Do not rely entirely upon heart rate meter alarm signals. Keep
pacemaker patients under close surveillance.
Pacemaker monitoring is not possible with ECG cables that have unshielded lead
wires. Ensure that only shielded lead wire ECG cables are used when monitoring
patients that have a pacemaker.
FORA recommends to apply a SpO
to set the alarm range for the peripheral pulse (PP) in the range of the heart rate (HR),
or to set the HR source in the SpO
Caution!
The guidelines in this section are for overview only. They are not a substitute for, nor
do they overrule manufacturers' documentation and instructions or departmental
procedures.
SpO
Warnings
2
These warnings and cautions pertain to the SpO
WARNING!
The SpO
measurements. Do not attempt to connect these cables to a PC or any other monitor.
Do not use a damaged probe or extension cables.
Incorrect application or extended use of a SpO
cause incorrect readings. Inspect the sensor site periodically as directed in the
probe and extension cables are intended for use only for pulse oximeter
2
sensor in addition to the ECG measurement and
2
menu to SpO
.
2
2
feature of this monitor.
2
probe on the same sensor site may
2
3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VSM100 Pro and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents