The device is intended for use in the electromagnetic environment specified in the following tables.
The user should ensure that it is used in such an environment.
Immunity Test
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range. Interference may occur in the vicinity of
equipment marked with the following symbol:
Information subject to change. Please visit
RF Electromagnetic immunity
Test Level
EN 60601-1-2
3 V
RMS
150 kHz to 80 MHz
6 V
in ISM and
RMS
amateur radio bands
between 150 kHz to
80 MHz
3 V/m
80 MHz to 2,7 GHz
3 V/m
(Proximity fields)
www.seco.com
Compliance Level
3 V
RMS
150 kHz to 80 MHz
6 V
in ISM and
RMS
amateur radio bands
between 150 kHz to 80
MHz
3 V/m
80 MHz to 2,7 GHz
3 V/m
(Proximity fields)
to find the latest version of this manual. Page.27 of 29
SYS-D14-MED
Electromagnetic
Environment Guidance
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the device,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended
separation distance
d = 1,2 ⋅√P 150kHz to
80MHz
d = 1,2 ⋅√P 80 MHz to
800MHz
d = 2,3 ⋅√P 800 MHz to
2,5GHz
where P is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and d is the
recommended separation
distance in meters (m).