To install the device properly on the destination machine, follow the procedure below:
Make sure the electric connection on the destination machine is off and the
1
electric power supply is disconnected from the device.
Proceed to fasten the device to the destination machine, using the relative
2
screws in points (A), indicated in the figure.
Proceed with the electrical connection of the device.
3
Proceed to electrically connect the destination machine.
4
For wallmount installation, follow the procedure below:
Make sure the electric connection on the destination machine is off and the
1
electric power supply is disconnected from the device.
Note: for wallmount configuration turn the device to keep the sink fins
2
upright. Proceed to fasten the device to the wall, using:
expansion screws
material: Nylon
o
type: SX5-SP 6X40mm
o
hole diameter: 6mm
o
o hole depth: 40mm
screws
o type: 4.5 x 60mm cylindrical head
plain washer
o internal diameter 4.3mm
o external diameter 16mm
o width 1.6mm
o compliant with ISO 7089 in
points (A), indicated in the figure.
Proceed with the electrical connection of the device.
3
Proceed to electrically connect the destination machine.
4