GE H
EALTHCARE
D
5180263-100, R
IRECTION
7-5-18-7
Network Adapter
See:
7-5-3 "Diagnostics Window Overview" on page 7-24
Runs a test on essential functions of the Network Interface. The test requires the system to be
connected to a network. The test results are displayed in the Status portion of the window.
This test gives the current status of the network and provides the option to restart the network
components in the BEP.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
Click the Execute button.
5
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
for a description of this window's features.