Download Print this page

GE ÄKTAprocess Assembly Instructions page 3

Air trap sightglass cover
Hide thumbs Also See for ÄKTAprocess:

Advertisement

28-9266-70 AA
4
Mount the bottom sightglass cover part
Slide the bottom cover (H) over the lower ends of the two
front connecting rods (F) under the support (D) and align
the corresponding holes on both cover parts.
Fasten the bottom cover with two nuts (E) and tighten
securely. Use a suitable wrench to hold the bottom nut (C)
steady while tightening, to avoid increasing the strain on
the connecting rod.
Note: Do not use washers on the front connecting rods.
Fasten the other connecting rods to the support (D).
5
Connect the cover parts
Fasten the two cover parts together using bolts (I),
washers (J) and nuts (K).
Tighten securely.
6
Final connections
Reconnect the TC coupling (A) and evacuation hose to the
top valve.
Attach the cable contacts (B) to the level sensors.
Assembly mount procedure – 1 inch air traps
1
Disconnect the evacuation hose and remove the
bottom nuts
Remove the TC coupling and evacuation hose from the top
valve (A).
Loosen and remove the bottom nuts (B) and washers from
the two front connecting rods.
Note: The washers from the two front connecting rods will not
be used when the cover is mounted.
2
Pull out the connecting rods
Pull the connecting rods (C) out of the air trap. Remove the
washers from the rods.
Note: Do not remove the dome nuts. Ensure that the dome nuts
still are securely fastened. If loose, the nuts should be
fastened and locked into place using Loctite™ 243.
Sightglass cover assembly instruction
H
I
J
A
B
C
E
3

Advertisement