Smiths Medical level 1 Operator's Manual page 29

Convective warmer
Hide thumbs Also See for level 1:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Emissions test
Compliance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Note: Compliance using 100-240V 50/60Hz with AC power cord of 3 m - 5 m, with hose thermistor cable of 2 m
WARNING:
The use of power cord and hose other than those listed in the electromagnetic emissions
declaration may result in increased emissions or decreased immunity of the convective warmer.
The convective warmer should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, you should verify normal operation of the convective
warmer in the configuration in which it is to be used.
Common portable and mobile consumer electronic devices may cause interference with the
convective warmer. Observe the convective warmer to verify normal operation.
Facility wiring must comply with all applicable electrical codes. Do not bypass power cord
connections. Do not remove a prong from the power cord.
Group 1
The convective warmer uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class A
Note: The EMISSIONS characteristics of this equipment make it suitable
for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a
Class A
residential environment (for which CISPR 11 class B is normally required)
this equipment might not offer adequate protection to radio-frequency
Complies
communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Level 1®
CONVECTIVE WARMER
S E C T I O N 8
Electromagnetic environment - guidance
| Operator's Manual
S p e c i f i c a t i o n s
31

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

L1-cw-100vL1-cw-120vL1-cw-220v

Table of Contents