S E C T I O N 7
T r o u b l e s h o o t i n g
Problem
All indicators are flashing and
the warmer will not run
Electromagnetic interference
to other devices
Electromagnetic interference
from other devices
28
Possible Cause/Solution/Operating Conditions
1. A button was pressed during power up
• Remove the power plug from the Mains, verify that no buttons are stuck or
pressed, re-insert the power plug and watch for the system to start normally
2. A system fault has occurred
• Remove the power plug from the Mains and then re-insert the power plug,
watch for the system to start normally
This equipment has been tested and found to comply with the limits for medical
devices to IEC 60601-1-2/EN 60601-1-2. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation.
This equipment generates, uses, and can radiate radio frequency energy. If it is not
installed and used in accordance with the instructions, it may cause harmful
interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference to other devices, which can be determined by removing power
(unplug power cord) and then reapplying power, the user is encouraged to try to
correct the interference by one or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment to an outlet on a circuit different from that to which
the other device(s) is (are) connected.
• Consult the manufacturer or field service technician for help.
If placed near monitoring equipment, or other electronic devices, this equipment
must be IEC 60601-1-2/EN 60601-1-2 approved or electromagnetic interference
may result.
Move the convective warmer away from the portable or mobile RF communications
equipment per the recommended separation distance specified in Section 8,
Specifications.