Van Walt Rain Gauge Operating Instructions Manual page 20

Table of Contents

Advertisement

VAN WALT
Monitoring your needs
• Replace the battery of the datalogger in time with a new battery of the correct type. Chapter
2, "Technical specifications," states the expected life span of the battery. For replacement
of the battery, a battery set is available separately (see Chapter 1.1, "Package contents").
Consult the user manual of the battery set e+ logger (11.41.90.01) to replace the battery
correctly.
If the datalogger does not function correctly after replacing the battery, short circuit the battery for
5 seconds and then end the short circuit in one direct movement.
The rain gauge is factory-calibrated and is supplied with a calibration specification, elucidated
with examples. The instrument is specified with a nominal measuring sensitivity of 0.2 mm rain per
contact. The actual value may lie between 0.197 and 0.204 mm per contact. In addition to the
nominal specification, the calibration specification states the actual measuring sensitivity of the
instrument, the so-called calibration factor. The user has the option of calibrating the instrument, to
which a static calibration is performed. Subsequently, the new calibration factor of the instrument
is determined via a dynamic calibration measurement. If so desired, the Range setting of the
datalogger can then be adjusted in accordance with the new calibration factor.
The static calibration is executed as follows:
• Preceding the static calibration, all maintenance requirements as stated above should be
completed.
• Remove the water-collecting basin and check if the rain gauge is mounted horizontally.
• With a pipette or burette, slowly drip 10.13 cm3 water in the tipping bucket. It should tip with
the last drop from the pipette or burette. If this does not occur, adjust the relevant thumbscrew
and repeat the procedure until the tipping bucket tips at the right moment.
• Repeat the above procedure for the other half of the tipping bucket.
Next, the calibration factor is determined according to the method for dynamic calibration:
• Before the calibration factor is determined, the rain gauge should be connected to the
datalogger, the water- collecting basin should be removed from the rain gauge, and a check
should be performed to see if the rain gauge is standing in the horizontal position.
Number of contacts
96.7 - 96.9
97.0 - 97.4
97.5 - 97.9
98.0 - 98.4
98.5 - 98.9
99.0 - 99.4
99.5 - 99.9
100 - 100.4
Van Walt Ltd | Prestwick Lane | Grayswood | Haslemere | Surrey | GU27 2DU | Tel. 01428 661 660 | Fax. 01428 656 808 | www.vanwalt.com
Operating Instructions
Rain Gauge
calibrationfactor
[mm per contactsluiting]
0.204
0.203
0.202
0.201
0.200
0.199
0.198
0.197
20

Advertisement

Table of Contents
loading

Table of Contents