Maintenance
90
Step
Display
8
DETERMINE EMPTY VOLUME DOSING
PUMP CONVEYS SAMPLE FOR INJECTION
WHEN FIRST DROP FALLS AT THE CAPILLARY
STOP PUMP!
[E] START PUMP
9
DETERMINE EMPTY VOLUME DOSING
HAS FIRST DROP FALLEN
[E] STOP PUMP
10
DETERMINE EMPTY VOLUME DOSING
EMPTY VOL. DOSING: xxx ul
[E] TO CONFIRM
11
Measuring operation is started once you confirm with the [E] key.
"Empty volume dosing" ends.
TOCII CA72TOC
Description
Start the pump with the [E] key.
Stop the pump with the [E] key when the
first drop falls.
The new volume value determined
appears on the display. The value,
including the volume and time, is entered
in the EMPTY VOLUME DOSING log.
The value determined is saved in the
EMPTY VOLUME P2 [μl] parameter.