Lamp Replacement and Calibration Corona extreme
EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
14
Changing the device profile
The counter must be reset each time the halogen lamp is replaced.
When a lamp hour reset is performed, all profiles stored before the lamp was
replaced are deactivated. Therefore it is necessary to calibrate all profiles on the
sensor as described in step 6.
Click Yes to start Reset.
Fig. 56 Resetting counter
Several steps now need to be taken to store the current reference as the new
default setting on the device.
Fig. 57 Saving new configurations
Performing sensor calibration
47