3. Operation

Turn the unit on using the main switch on the frontal panel.
The display shows the software version and the last setpoint value. Automatically, the unit starts the thermoregulation to
achieve and maintain the set temperature.
The internal temperature is measured by the internal probe and it is shown on the display.
To adjust the working temperature, press the "Set" button. Then use the "+" and "-" buttons to increase or decrease the
setpoint value, respectively (by keeping the buttons pressed, the setpoint can be adjusted faster).
The new setpoint is confirmed by pressing again the "Set" button or automatically if no buttons are pressed.
Use the dedicated switch on the frontal panel to control the internal sockets.
Placing products / samples in the incubator
Do not place hot products in the incubator; otherwise, the average internal temperature will raise and the other products
inside can be damaged.
Do not often open the door to avoid big internal temperature variations and to reduce energy consumption.
To ensure good air circulation within the unit, avoid placing large quantity of products and do not place objects against the
inside back panel.
Do not place any product too close to the temperature probe.
Connect only 230 V (current consumption 3 A max.) instruments to the internal sockets.
1
Main switch
2
Internal sockets switch
3
Wi-Fi led
4
Display
5
Control buttons