4. Setting mode

Press the "Set" button for 3 seconds to access the setting mode.
Use the "+" and "-" buttons to browse through the available menus, listed below.
Press the "Set" button to access the selected menu. Then, use the "+" and "-" buttons to modify the value of the parameter
(when it is possible).
Press the "Set" button to confirm.
To exit the setting mode, do not press any buttons for 5 seconds.
Text shown on
the display
CAL
ULI
LLI
CCO
CHE
CFA
UFI
OtC
AP
rSt

5. Error and warning messages

When the display shows an error message, the instrument's functions are stopped automatically.
AL1
AL2
AL3
AL4
AL5
To remove the error message, restart the instrument.
If the alarm persists on the display, please contact VELP Scientifica's technical service department.
Default
Range
value
- 9,9 ÷ 9,9 °C
- 0.5 °C
0,1 °C steps
LLI ÷ 50 °C
50 °C
0,1 °C steps
3 ÷ ULI °C
3 °C
0,1 °C steps
---
YES – NO
NO
---
YES – NO
NO
YES – NO
NO
Internal temperature too low
Internal temperature too high
Internal probe broken
Time to reach setpoint too high
Big internal temperature variation
Menu's name and description
Probe calibration
It allows to align the temperature probe reading to a
reference thermometer
Setpoint upper limit
It limits the maximum value of temperature setpoint
Setpoint lower limit
It limits the minimum value of temperature setpoint
Compressor, heating element and fan operating time
It shows compressor, heating element and fan operating
times.
Operating times are showed in hours up to 999 hours
(around 42 days). Then they are showed in days using
different combinations of the digits' decimal point.
Wi-Fi connection
It allows to enable / disable the Wi-Fi connection
One time code
It shows the One Time Code during the registration of
the device on the VELP Ermes cloud platform
Access point
It allows to enable / disable the Access Point mode of
the Wi-Fi connection
Reset
It allows to reset all the system settings to the default
values
Warning messages only on Ermes platform