High Voltage Probe HV-P60A
● Measure in accordance with operating procedure when using
this instrument.
Erroneous operating procedure (refer to "3. Adjustment of Probe" and "4.
Measuring method".) may result in an electric shock, fire or malfunction.
● When connecting the object to be measured to the nose of the
probe, firmly fix a measuring lead with an accessory Clamp.
Imcomplete clamping may result in an electric shock, fire or malfunction.
● Do not bring a grouding lead into contact with an insulator or
the nose of the probe.
If the grouding lead is brought into contact with the insulator or nose of the
probe, it may result in an electric shock, fire or malfunction.
● Coating of the grouding lead is voltage resistant (approx. 10
kV, but a clip is not and should be used carefully.
● Do not apply a voltage higher than the specified one.
If a voltage higher than the specified is applied, it may result in an electric
shock, fire or malfunction. The maximum input voltage is 60 kV at DC to 20
kHz. If the frequency exceeds 20 kHz, the maximum input voltage will
decrease (see "Fig.7 Maximum Input Voltage vs. Frequency").
● During or before measurement, make sure that no one or
nothing come into contact around you.
If you touch someone or something during measurement, it may result in an
electric shock.
Read the following safety information
Warnings
[Read the next page.]