These instructions describe how to replace the IR Optical Bench in the AGM300 Ammonia Gas Monitor.
It is assumed that the user is familiar with the operation of the AGM300 / ADM800 Ammonia Gas
Monitoring System. If necessary, refer to Instruction 3015-4275 for detailed operation and maintenance
information.
Items Required:
Replacement IR Optical Bench Kit 3015-4521:
IR Bench Assembly 3015-4204
o
Wire Ties (Qty. 2) 3015-2835
o
Instruction Sheet 3015-4723
o
Medium Phillips Head Screwdriver
Small Flat Blade Screwdriver
Wire Cutter (for removing wire ties)
Removing the IR Optical Bench
Figure 1. AGM300
Ammonia Gas Monitor
Replacement Kit 3015-4521
1. Find the AGM300 Ammonia Gas Monitor and switch OFF its AC
circuit breaker. Observe that the monitor's green MONITOR ON
indicator light on the front panel should be OFF.
2. Use a Phillips screwdriver to loosen the monitor's two door latches,
and then open the door.
3. Take a few minutes to familiarize yourself with the internal
components of the AGM300 shown in Figure 2.
Figure 2. AGM300 Components
AGM300 Optical Bench
Instruction 3015-4723
Upper Gas Sample Tube
Upper Electrical Connector (Red)
Upper Wire Tie
Main Board
Lower Wire Tie
AC Power Switch
Lower Electrical Connector
Lower Gas Sample Tube
Rev. 0 – February 2006

Advertisement

loading

Summary of Contents for Bacharach AGM300

  • Page 1 These instructions describe how to replace the IR Optical Bench in the AGM300 Ammonia Gas Monitor. It is assumed that the user is familiar with the operation of the AGM300 / ADM800 Ammonia Gas Monitoring System. If necessary, refer to Instruction 3015-4275 for detailed operation and maintenance information.
  • Page 2: Installing The New Ir Optical Bench

    Using wire cutters, cut the upper and lower wire ties that hold the IR Optical Bench in place, and then remove the bench. d. Please provide the serial number of the AGM300 and return the old IR Optical Bench to: Bacharach Inc.
  • Page 3 10. Check, and possibly adjust, the IR Optical Bench detector voltage by first going to the ADM800 and pushing the button on the System Screen that corresponds to the AGM300 monitor being serviced. This action will display that monitor’s SETUP Screen.
  • Page 4 Website: www.bacharach-inc.com • E-mail: help@bacharach-inc.com Printed in U.S.A. 13. Go back to the AGM300 and find GAIN ADJ potentiometer R34 on the Main board. Using a small flat-blade screwdriver, turn R34 clockwise to increase, or counterclockwise to decrease, the detector voltage to between 4.2 and 4.3 VDC.

This manual is also suitable for:

3015-4521