7. Maintenance

7.1
General
The CuroCell® IQ Cirrus hybrid mattress is a medical
device, in accordance with Medical Device Regulation
MDR (2017/745), that is used as an aid in preventing and
treating pressure ulcers/pressure injuries. For this reason,
we recommend that the control unit be regularly serviced
and inspected to maintain functionality and performance.
Like other technical devices, a control unit that is properly
cared for works better and lasts longer.
Servicing and maintenance must always be performed by
Care of Sweden or one of its authorized technicians. Only
use spare parts that have been approved by Care of
Sweden. Use of non-approved spare parts invalidates the
warranty. For more information, see the CuroCell® IQ
Cirrus service manual.
Faults and defects regarded as warranty faults are fixed
free of charge.
A warranty claim must be submitted to Care of Sweden be-
fore the product is returned
After the product is used, check that:
1.
The power cable and power supply are undamaged.
2.
The connecting tubes (marked CPR) on the side of the
control unit are positioned correctly and not leaking.
3.
The hygiene cover is intact and the cover and cells are
correctly assembled.
4.
No tubes or connectors are damaged or jammed.
Contact Care of Sweden or your local distributor if any spare
parts are required.
7.2
Replacing the air filter
Before any maintenances are done, make sure that the
control unit is turned off. Services shall not be done while
using the product.
To replace the air filter:
1. Loosen the small protective plate on the rear of the
control unit using a size T10 Torx screwdriver.
2. Remove the filter from the holder.
3. Place the new filter in the holder with the pink side facing
outwards. Put the protective plate back in place and secure
using the screws.
If the control unit is used in a very dirty environment the
filter should be checked regularly.