Reuse, Cleaning And Reconditioning; Storage - Care of Sweden CuroCell IQ Cirrus Instructions For Use Manual

Automatic active airflow controlled mattress system
Hide thumbs Also See for CuroCell IQ Cirrus:
Table of Contents

Advertisement

5. Reuse, cleaning and reconditioning

The product is reusable. When reusing, it is important to
follow the instructions below for cleaning and
reconditioning. Before using the product again for a new
user, it is recommended to clean the mattress in
accordance with the instructions below. Consult your
hygiene manager or Care of Sweden for help and
instructions if you are unsure.
Cleaning and reconditioning should otherwise be performed
as required.
5.1
Cleaning
CONTROL UNIT
Wipe off the control unit with a damp cloth and
mild cleaning solution for this purpose: such as
alcohol with or without tenside surfactants,
oxidizing solutions, chlorine max 1 % or hydrogen
peroxide max 1,5 %.
If using a different agent, choose one that does not affect the
control unit.
MATTRESS COVER
Covers can be wiped off with a mild cleaning solu-
tion for this purpose such as: alcohol with or with-
out tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % (max 10% for cover Stone) or
hydrogen peroxide max 1,5 %.
Excrement and blood stains must be removed as soon as
possible using cold water. Carefully follow local instructions
and the instructions for the detergent.
Multi-layer covers should be divided before washing. The
parts should be washed with similar colours.
Chlorine and phenol-based cleaners could adversely
affect the PU surface and should be avoided. If chlo-
rine is used, we recommend a mixture of max 1 %
(max 10% for cover Stone).
INNER COVER
Clean the affected area with a mild cleaning solution for this
purpose such as: alcohol with or without tenside surfactants,
Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or
hydrogen peroxide max 1,5 %.

6. Storage

It is advisable to store the mattress and control unit in the
product bag (accessory), original package or equivalent for
protective storage. Handle the packaged product with
caution. Do not place any heavy objects on top of it. For
additional information about storage temperature, see
section 9.
5.2 Reconditioning
CONTROL UNIT
See section 5.1 for instructions.
MATTRESS
1.
Disconnect the tube connector from the control unit
and remove the air from the mattress.
2.
Remove the covers. For more information of cleaning
the covers, see section 5.1.
3.
Move the mattress to a clean area that is suitable for
cleaning.
4.
Wipe off the cells, all tubes and the CPR module with a
mild cleaning solution for this purpose such as: alcohol with
or without tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen peroxide max 1,5
%. Let it dry.
5.
Wipe off the work surfaces using a suitable detergent and/
or disinfectant.
6.
Place the air cells back into the foam core. If cells for any
reason have become loose from the tubes, these must be
replaced according to the drawing in section 4.2.
FOAM CORE
Clean the affected area with a mild detergent (such as
washing-up liquid) and water or with an alcohol-based
disinfectant
(cleaner
Gently squeeze out any water.
NOTE:
Check the hygiene cover, cells and hoses each time the product is
cleaned.If damaged, it must be replaced or repaired. Also check
the control unit, tube connectors and power cable during
cleaning. Damaged parts must be replaced or repaired.
Do not wring or roll the foam core to extract the water. Let it
dry in a warm, ventilated area (not in direct sunlight). The foam
core must be completely dry before it is used again.
14
intended
for
this
purpose).