3.6.1 Table of notifications
Information about each notification is shown in the
notification table:
Notifications from 1-10 are both audible and visual.
The error code will be displayed until the error has been
rectified. If the mute button is pressed the audible
warning will crease for a period of 5 minutes and will
return until the error has been rectified.
The notifications 11-13 have no audible alarm. The error
code is shown until the system is restarted.
Notification
1
2
3
4
5
6
7
8
9
Description and
troubleshooting
High temperature. Valves and
compressors are turned off.
If the control unit is in direct
sunlight, relocate it.
Otherwise, contact the
support.
Default settings are not
completed. Contact the
support.
Incorrect input voltage. Make
sure that the correct power
supply is used, otherwise
contact the support.
Low pressure. Secure the
CPR, mattress, air tubes and
air filter. If the problem
remains, contact the support.
Automatic setting failure. The
correct pressure has not been
reached within the time limit.
If the problem remains,
contact the support.
Leakage under the automatic
setting period. The mattress
is leaking too much for the
weighing to be accomplished.
Control the mattress and the
air connections. If the
problem remains, contact the
support.
High pressure. The pressure
cannot be reduced to the
desired value within the time
limit. Contact the support.
The automatic setting has
been restarted too many
times during the automatic
setting period. Contact the
support.
The mattress control
parameters have not been
read. Connect the CPR or
contact the support.
10
11
12
13
3.6.2 Maximum pressure notification
When the function Maximum pressure has
been used for a long time, the Maximum
pressure diode will blink. If the use is
intentional, ignore the notification.
3.7 Transport function
If the user needs to be moved in bed, there are two ways
to perform the transportation:
Unplug the CPR connection, close the lid, place the CPR
connection at the end of the bed and remove the
control unit from the bed. The mattress will retain air
for at least 12 hours.
or
Remove the power supply from the wall socket and
leave the control unit hanging on the bed during
transport. The mattress will retain air for at least 12
hours.
We recommend using this function for short periods only.
10
The mattress control
parameters have been
changed during the use.
Restart the system. If the
problem remains, contact the
support.
Leakage in one of the cell
sections. Secure the CPR,
mattress and connection
tubes. If the problem
remains, contact the support.
Leakage in one of the cell
sections. Secure the CPR,
mattress and connection
tubes. If the problem
remains, contact the support.
Leakage in one of the cell
sections. Secure the CPR,
mattress and connection
tubes. If the problem
remains, contact the support.