1.6

Internal detector tests

The detector performs two types of self-tests:
Continuous feature test
Built-in test (BIT)
1.6.1
Continuous feature test
The detector is supplied with default settings, containing Continuous Feature Test. To
change these settings, refer to Modbus
During normal operation, the detector tests itself continuously and indicates a fault if a
failure is found. This type of test complies with SIL-2 requirements.
The detector continuously tests:
Input voltage level
All internal regulator voltage level
Voltage level status of sensor and sensor circuitry for noise or disconnection in the
electronic circuitry
0-20 mA level output
Relays and heater operation
Processor watch dog
Software
Memory
Oscillator frequency
Response to a fault indication
If a failure is found, the detector indicates it by:
Fault relay:
0-20 mA: indicates fault (0 mA or 2 mA) in output configurations 1A, 2A, and 3A
LED - yellow flashes (4 Hz)
Correcting the fault
Rosemount 975UF
Opens in output configurations 1A, 2A, and 1R
Closes in output configurations 3A and 2R
®