Interchangeable Configuration - AGFA CR User Manual

User station
Table of Contents

Advertisement

CR U
S
SER
TATION

Interchangeable configuration

In an interchangeable configuration one or more CR User Stations can
serve a range of Digitizers.
In this configuration, a cassette can be identified with any of the
CR User Stations and can be scanned by using any of the Digitizers. The
patient
demographic
CR User Station and stored in the memory chip of the cassette. As a result
the identification data are linked to the cassette, and any Digitizer can be
used to scan the cassette.
This configuration allows the flexible use of several Digitizers and
CR User Stations, depending on the workload.
CR User Station(s)
with ID Tablet
The transmission of the identification data to the cassette chip takes place
when you insert the cassette into the transmission slot of the
CR User Station's ID Tablet. When the cassette is read in the Digitizer, the
ADC Quality System links the demographic data to the image. The study
data are used to set the image processing parameters.
10
Introducing the CR User Station
data
Digitizer(s)
and
study
data
Interchangeable Configuration
are
entered
via
2313B EN 20041109
the