TABLE OF CONTENTS WARRANTY ......................1 FOREWORD ......................2 SYMBOLS GLOSSARY ..................2 PRECAUTIONARY INSTRUCTIONS ............3 INTENDED USE .....................5 DEVICE DESCRIPTION ..................6 UNPACKING AND INSPECTION ..............8 INSTALLATION ......................9 OPERATING INSTRUCTIONS ................10 CARE & MAINTENANCE .................19 TROUBLESHOOTING ..................20 ELECTRICAL INTERFERENCE ..............21 APPENDIX A - EMC TABLES.................21 WARNING (USA ONLY): THIS DEVICE SHOULD BE USED ONLY UNDER THE CONTINUED SUPERVISION OF A PHYSICIAN, LICENSED PRACTITIONER OR...
This warranty gives you specific legal rights and Richmar APT Performer. Richmar warrants that you may also have other rights which vary from your APT Performer is free of defects in material location to location. Richmar does not authorize and workmanship. This warranty shall remain...
FOREWORD This manual has been written for the as the accessories before operating the users of the Richmar APT Performer. It system. It is intended for use in treatment contains general information on operation, facilities and rehabilitation clinics to offer...
• The APT Performer should be kept out of • Before starting leg training, the arm trainer the reach of children. should be swiveled so that the handle •...
Page 6
• Before plugging in the APT Performer into • Please avoid the excessive vibration of the a main socket, ensure that the voltage of device.
DO NOT use the APT Performer specified heart rate sensor to be used with simultaneously with other therapeutic device the APT Performer should be used no closer (such as shortwave), to avoid mis-operation. than 30 cm (12 inches) to any part of the APT Performer, including cables specified by •...
CONTRAINDICATIONS feels is not suitable to use this device. Patients with the following conditions should not use the APT Performer: ADVERSE EFFECTS Ligaments, knees or joint ruptures Discontinue use if the patient experiences adverse reactions from the device. Possible...
Page 9
DEVICE DESCRIPTION TECHNICAL SPECIFICATIONS FOR DQAPT GENERAL Dimensions (W×L×H) 28" × 30" × 42" Standard Weight 102 lbs Interface 8" touch screen Expected Useful Life Time 5 years Software Release Version SAFE WORKING LOAD UE Pedals (total) 11 lbs LE Pedals (total) 22 lbs ADJUSTABLE RANGE Supporting module...
Page 10
DEVICE DESCRIPTION BLUETOOTH® The Bluetooth® word mark and logos are Communication between the device and owned by the Bluetooth SIG, Inc. and any the heart rate sensor (not included) is via use of such marks by Compass Health Bluetooth. Brands, Corp. is under license. TECHNICAL SPECIFICATIONS (CONTINUED) BLUETOOTH Bluetooth Version...
The system will boot into self- WARNING inspection mode, as shown below. Before you plug the APT Performer into a main socket please check that the voltage of the device stated on the marking corresponds with the voltage of the power supply.
OPERATING INSTRUCTIONS TOUCHSCREEN DISPLAY - NAVIGATION BAR ITEM DESCRIPTION Indicates that the device has connected with the Bluetooth Heart Rate Sensor. Indicates that the USB flash disk has been inserted into the device and recognized. See Training Date Export Section for more details.
Page 13
OPERATING INSTRUCTIONS 3. Adjusting the direction of arm trainer CAUTION Before Arm training, the arm trainer can be Before each leg training, check the be adjusted to 180° using the knobs next patients feet to ensure they are secure and to the handlebar as seen below.
Page 14
NOTE: The Date and Time setup is patient’s treatment. Otherwise, press necessary for the APT Performer to desired training program to move to the properly export training data. If the next screen.
Page 15
OPERATING INSTRUCTIONS Spasm Control GAMEPLAY When the patient has a spasm during Each game interface shows tips about how training, an indication will appear on the to play each game. Below is one example. screen along with an audible warning Game Status and Operation Bar sound, signaling spasm protection has activated.
Page 16
OPERATING INSTRUCTIONS Adjusting the Training Parameters during speed, the system will automatically switch to the ACTIVE mode with a voice prompt. Training When it is detected that the current speed If you need to change the training of the motor is lower than 10 r/min of parameters during training, please press the set speed, the system will return to the PAUSE button.
Page 17
OPERATING INSTRUCTIONS Endurance Training Once the parameters have been set, the game will display a tip on how to play the Endurance Training is an Active/Passive game, press OK to begin training. program. During Endurance training, the patient is motivated to meet the set Symmetry training was designed to mileage.
Page 18
OPERATING INSTRUCTIONS ROM Training Read the game tips and tell the patient how to train, then press OK button to start Range of Motion (ROM) Training is a training. Passive program designed to use the motor of the trainer to help the patient relearn how to maintain a fluid, constant motion increasing their range of movement in their limbs.
Page 19
OPERATING INSTRUCTIONS Training Data Summary STORING AND RECALLING FAVORITES Once training ends, the system will Storing Favorites automatically display the training To store a training session as a favorite summary, as shown in the example below. to use for future trainings or on other patients, press FAVORITES button from the Training Summary, then type in the name of the favorite program.
Page 20
ARM and LEG trainer of the device. Emergency Stop The APT Performer is equipped with an End of Training Emergency Stop function. At the end of training the trainer will beep and stop after 3 beeps.
The device and accessories must be checked will occur to the main device such as at regular intervals. deformation and chipping of paint. Before you plug the APT Performer • DO NOT use bleach to clean the straps. into a main socket please check that •...
TROUBLESHOOTING TROUBLESHOOTING ERROR POSSIBLE CAUSE SOLUTION No response after starting Lack of power. Check whether the the device. power cord is plugged Fuse might be into the device as well burned out. as the wall outlet. Contact Manufacturer for an authorized technician to replace the fuse.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS The APT Performer is intended for use in the electromagnetic environment specified below. The customer of the user of the APT Performer should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
Page 24
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY The APT Performer is intended for use in the electromagnetic environment specified below. The customer or the user of APT Performer should assure that it is used in such an environment. Immunity test...
Page 25
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY The APT Performer is intended for use in the electromagnetic environment specified below. The customer or the user of APT Performer should assure that it is used in such an environment. Electromagnetic environment –...
Page 26
APPENDIX A - EMC TABLES TABLE: TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATIONS EQUIPMENT Band Test Immunity Maximum frequency (MHz) a) Service a) Modulation b) Distance (m) TEST LEVEL power (W) (MHz) (V/m) Pulse 380 to 390 TETRA 400 Modulation b): 18Hz...