14 Exposure value tables for Planmeca ProX

14.1 Default exposure values

14.2 Recommended exposure values for animals with Planmeca ProX

Cats or smaller animals
Incisors
Premolars
Molars
Occlusal
Needle in
root canal
Bitewing
User's manual
The following table shows the default exposure values for Planmeca
ProSensor with no target selected.
Short cone
kV
Large
63
Small
60
Default exposure values described in A.1 are factory settings, by changing
only the time of these default exposure values you can easily set and save
the recommended exposure values for animals.
For detailed instructions on how to program new exposure values for each
ProX exposure region, see Planmeca ProX for veterinary use user's manual.
kV
mA s
Upper jaw
60
8
Lower jaw
60
8
Upper jaw
60
8
Lower jaw
60
8
Upper jaw
60
8
Lower jaw
60
8
Upper jaw
66
8
Lower jaw
66
8
Depends on
60
8
teeth
Not used for animals
14 Exposure value tables for Planmeca ProX
mA
s
8
0,1
8
0,08
Larger animals
0.100
Incisors
0.100
0.100
Premolars
0.100
0.100 –
Molars
0.125
0.100 –
0.125
0.125
Occlusal
0.100
0.100 -
Needle in
0.125
root canal
Bitewing
Long cone
kV
mA
Large
63
8
Small
60
8
kV
mA s
Upper jaw
60
8
Lower jaw
60
8
Upper jaw
63
8
Lower jaw
63
8
Upper jaw
63
8
Lower jaw
63
8
Upper jaw
70
8
Lower jaw
70
8
Depends on
60
8
teeth
Not used for animals
Planmeca ProSensor® HD
s
0,2
0,16
0.160
0.160
0.200
0.200
0.250
0.250
0.250
0.200
0.160 -
0.250
23