5 First Operating Steps | 5.7 Adjusting the Transmitted-Light Phase Contrast or the Transmitted-Light
Darkfield using the Turret Condenser
5.7 Adjusting the Transmitted-Light Phase Contrast or the Transmitted-Light
Darkfield using the Turret Condenser
Prerequisite
ü
The phase contrast method on the microscope requires a stand with Full-Köhler equipment..
ü
The turret condenser with phase contrast stop or darkfield stop inserted is required.
Procedure
1. Adjust the microscope as you would for transmitted-light brightfield, see Adjusting the
Transmitted-Light Brightfield on the Full-Köhler Microscope [
2. Turn the nosepiece to move the phase contrast
objective
path.
3. Open the luminous-field diaphragm on the
stand.
4. Turn the turret disk of the turret condenser to
the phase contrast position
Ph 3) that corresponds to the objective being
used.
78
(Ph 1
or
Ph 2
or Ph 3) into the light
(Ph 1
or
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en
}
 69].
Ph 2
or
ZEISS