Introduction
The Type 1 91 Constant Ampl itude Signal Generator is
a stable i nstrument which wi l l provide many hours of
trouble-free operation.
However, to i nsure measurement
accuracy, it is suggested that you recali brate the instru
ment after each 500 hours of operation, or every six months
if used intermittently.
It wi ll also be necessary to recali
brate certai n sections of the i nstrument when tubes, tra n
sistors or other components that affect the cal ibration
accuracy of the i nstrument are replaced.
This cali bration proced ure can be used either for com
plete calibration of the Type 1 91 to return it to original
performance, or as an operational check of instrument per
forma nce.
Completion of every step in this procedure
returns the Type 1 91 to origi nal factory performance stand
ards. If it is desired to merely touch up the calibration,
perform only those steps entitled Adjust.
NOTE
The Adjust steps p rovide a check of i nstrume n t
p e rfo rm a n c e before t h e a d justme n t i s m a d e . To
preve n t
t h e
n e ed
for
circuits w h e n p e rformi n g a partial ca l ibration,
readjust only if the l isted tolerance is not met.
General Information
Any needed maintenance should be performed before
proceeding with cali bration. Troubles which become a ppar
ent during calibration should be corrected using the tech
niques given in the Maintenance section of the Instruction
Manual.
This procedure is a rranged in a sequence which allows
this i nstrument to be calibrated with the least interaction
of adj ustments and reconnection of equipment.
the steps may be performed out of sequence or a step may
be done i ndividually.
However, some adjustments affect
the cali bration of other circuits within the instrument. In
this case, it will be necessary to check the operation of
other parts of the instrument. When a step i nteracts with
others, the steps which need to be checked will be noted.
The location of test points and adj ustments is shown i n
each step. Waveforms which a r e hel pful in determi ning the
correct adj ustment or operation a re also shown.
EQUIPMENT REQUIRED
( se e Figs. 6-1 a n d 6-2 )
General
The following equipment or its equivalent is required
for complete calibration of the Type 1 9 1 . Specifications
given a re the minimum necessary for accurate ca li bration
of this instrument. All test eq ui pment is assumed to be cor
rectly calibrated and operating withi n the original specifi-
®
SECTION 6
CAI.IBRATION
reca l ibration
of
other
If desired,
cations. I f equipment is substituted, it m ust meet or exceed
the specifications of the recommended equipment.
Special Test Equipment
For the quickest and most accurate calibration, special
ca l ibration fixtures a re used where necessary. All calibra
tion fixtures
listed
under
obtained from Tektronix, I nc. Order by part number through
your local Tektronix Field Office or representative.
1 . DC voltmeter. Mi nimum sensitivity, 20,000 ohms/volt;
accuracy, checked to within 1 "/0 at 6.2 and 1 1 volts, and
checked to within 3 "/0 at 450 volts. For example, Simpson
Model 262.
2. Real-time test osci lloscope. Band pass, DC to 1 0 MHz;
minimum deflection factor, 0.005 volts/division. Tektronix
Type 547 Oscilloscope with Type 1 Al p l ug-in unit or equiv
alent.
3. Sampling test oscil loscope. Bandwidth, equivalent to
DC to 1 000 MHz; minimum deflection factor, S mil livolts/
division. Tektronix Type 661 Oscilloscope with a Type 4S3
vertical plug-in and a Type 5T3 timing pl ug-in.
4. Standard amplitude calibrator.
within + 0.25 "/0 ; sig nal ampl itude, 0.01 volts to 2 volts;
output frequency, 1 kHz. Tektronix cali bration fixture 067-
0502-00 or equivalent.
5. Time-mark generator. Marker output of 1 j.tS, 5
1 0 j.tS; sine wave outputs of 20 ns a nd 50 ns; accuracy,
0.001 "/0. Tektronix Type 1 84 or equiva lent.
6. Variable autotransformer. Must be capable of sup
plying at least 25 watts to the Type 1 91 over a voltage
range of 94 to 1 38 volts ( 1 88 to 276 volts for 230-volt nom
i nal line).
If autotransformer does not have an AC volt
meter to i ndicate output voltage, monitor output with an
AC voltmeter (rms) with range of at least 1 38 (or 276) volts.
For exa m ple, General Radio Wl 0MT3W Metered Variac
Autotransformer.
7. Precision DC voltmeter.
meter resolution, 5 0 j.tV; range, 1 1 volts. For example, Fluke
Model 801 B.
8. 50 ohm voltage pickoff unit. Tektronix VP-2 or equiv
alent.
9. Termi nation.
I m peda nce 50 ohms; accuracy, +1 "/0 ;
connectors, B N C jack t o G R.
0083-00.
1 0. 1 00 X sampling probe. Tektronix P6035 or equiva
lent.
1 1 . 1 X probe. Tektronix P6028 or eq uiva lent.
1 2. Cable.
5 ns RG-8A / U cable with GR connectors.
Tektronix Part No. 01 7-0502-00.
1 3. Two cables.
Impedance 50 ohms; type RG-58A/U:
length, 42 inches, connectors, BNC.
01 2-0057-00.
Eq uipment Required
can be
A m plitude accuracy,
j.tS
and
Accuracy, within +0.05"/0 ;
Tektronix Part No. 01 7-
Tektronix Part No.
6-1