Environmental Conditions

The EL30000 Series is designed for indoor use. The table below shows the general environmental requirements for
this instrument.
Environmental condition
Temperature
Humidity
Altitude
Pollution degree
Installation category
Mains voltage (rms)
Mains frequency
Acoustic noise declaration
14
Requirement
Operating condition: 0 °C to 40 °C
Storage condition: –40 °C to 70 °C
Operating condition: Up to 85% RH at temperature up to 40 °C (non-condensing)
Storage condition: Up to 90% RH (non-condensing)
Up to 2000 m
2
II (for AC input)
100 VAC to 240 VAC ±10%
50/60 Hz
Sound pressure: Lρ < 65 dB(A) at operator position, Lρ < 70 dB(A) at bystander position
Sound power: Lw < 70 dB(A)