Specifying the Operation Mode (EL34243A only)

The Operation Mode allows you to specify the input capability when two independent inputs are being used.
– Independant : This operation supply two independent inputs capability to a maximum rating of +150 V, +60 A.
In Independent mode, the instrument is at its default input state.
– Parallel : This operation increases the current capability to a maximum rating of +150 V, +120 A. In Input-Par-
allel mode, the instrument doubles the input current on Input 1.
1. Press Input Settings > Operation Mode .
2. Press Independant or Parallel to select the desired operation mode.
3. Press Back to exit.
For input connection in Input-Parallel mode, see
From the remote interface:
To enable Input-parallel mode:
INP:PAIR PAR
78
Parallel Connections
for details.