6.12 70 – AdvancedUserCalibration
This command is used to execute a user calibration.
Command:
Command string:
Return string:
Parameter:
AngleBinary
Cal2Std
CalValue
Deviation
Example
Send
Receive
6.13 78 – AcceptUserCalibration
This command is used to accept the executed AdvancedUserCalibration. This valid after user
calibration or an accept command. If another command is executed the user calibration is not valid
anymore.
Command:
Command string:
Return string:
Parameter:
AngleBinary
Example
Send
Receive
10. August 2020
70
70|TID|AngleBinary|Cal2Std|CalValue:
70|TID|Deviation:
A single angle can be selected
0 = calibrate to working standard, 1 = calibrate on second standard
Calibration value, units depends on setting, if Cal2Std = 0 is CalValue = 0
Deviation in ppm (e.g. 6235 = 0.6235%) relative to the used sample
70|xy|1|0|0:
70|xy|2020:
78
78|TID|AngleBinary:
78|TID:
A single angle which was user calibrated before.
78|xy|2:
78|xy:
© Copyright 2020, PROCEQ SA
Command: 70
TID: xy
AngleBinary: 1 (angle A0)
Cal2Std: 0 (working standard)
CalValue: 0
Command: 70
TID: xy
Deviation: 2020ppm (0.2020%)
Command: 78
TID: xy
AngleBinary: 2 (angle A1)
Command: 78
TID: xy
18