In this section the command strings supported by the glossmeter are listed. Some possible access
by commands not described is provided for other use than inline (e.g. factory
configuration/calibration/diagnostic).
6.1 2 – AdvancedMeasureValue
This command executes a single measurement.
Command:
Command string:
Return string:
Parameter:
AngleBinary
ValueAn
Unit
Example
Send
Receive
6.2 3 – StartScan Measurement
This command starts the scan measurements until the device receives the stop command. Scan
measurement mode executes measurements at the fastest possible rate and sends them
consecuteively. Do not send other commands while scan mode is active. If you want to sent another
command, first execute the StopScanMeasurement command.
Command:
Command string:
Return string:
Parameter:
AngleBinary
ValueAn
Unit
Example
Send
10. August 2020
2
2|TID|AngleBinary:
2|TID|AngleBinary|Unit|ValueAn:
Angle combination
Measured value per angle n (n = 0...2), format %4.1f
GU or %
2|xy|3:
2|xy|3|GU|91.2|94.5:
3
3|TID|AngleBinary:
3|TID|AngleBinary|Unit|ValueAn:
Angle combination
Measured value per angle n (n = 0...2), format %4.1f
GU or %
3|xy|2:
© Copyright 2020, PROCEQ SA
Command: 2
TID: xy
AngleBinary: 3 (angle A0 and A1)
Command: 2
TID: xy
AngleBinary: 3 (angle A0 and A1)
Uint: GU
Measured values: A0 = 91.2, A2 = 94.5
Command: 2
TID: xy
AngleBinary: 2 (angle A1)