Set Medium Selection; Set Relay Mode - Endress+Hauser t-switch ATT11 Manual

Flow limit switch for liquids and gases
Table of Contents

Advertisement

Set-up Procedure

Set Medium Selection

This function is used to select a calibration curve for either liquid
or gas.
This operation is independent of the position of DIL8.
DIL 5
OFF
OFF
ON
ON
If a display is fitted it will show F1 for 3 seconds followed by the
selected curve for 3 seconds after any change of DIL 5-6.

Set Relay Mode

This function is used to configure the relay action at the
switching point.
This operation is independent of the position of DIL8.
DIL 7
OFF
ON
De-Energise at setpoint
If a display is fitted it will show either EnEr or dEEn for 3 seconds
after changing DIL 7.
21
DIL 6
Selection
OFF
L1
ON
L3
OFF
A2
ON
A3
Mode
Energise at setpoint
Flat Face
X (liquids only)
DO NOT USE DO NOT USE
DO NOT USE DO NOT USE
X (gas only)
Probe

Advertisement

Table of Contents
loading

Table of Contents