Operating manual for the TG uni 1, TG euro 1 and TG euro 1 med appliance tester
Conducting tests in accordance with VDE 0751-1/EN 62353, Example 4 - Devices with protective conductor (PC I) and applied part of type B
Entering the appliance identification
Press the [Appliance testing] button in the Main menu (Fig. 93).
A screen is displayed requesting entry of the appliance identification.
Pressing the [1/A] button changes the display to a keypad (Fig. 94)
allowing the entry of a test object ID number with a maximum of 19 digits.
You can also read this in using a barcode scanner (Fig. 95).
Entry of the appliance identification number is not compulsory.
Fig. 94
Fig. 95
The subsequent entry of the appliance type (Fig. 96) is not compulsory and
is automatically displayed for barcode entry. For manual entry you use the
[1/A] and [OK] buttons to reach this menu.
Press the [OK] button if you do not wish to enter any information here.
In the Main menu (Fig. 97) press the [to DIN VDE 0751-1] button.
Fig. 96
56
Fig. 97