Operating manual for the TG uni 1, TG euro 1 and TG euro 1 med appliance tester
Conducting tests in accordance with DIN VDE 0701-0702, Example 2 – Devices without a protective conductor (PC II)
Test settings
Fig. 47 shows the Main menu for test object protective measures. After
selecting the protective measures by pressing the [device w/o PE (Pc II)]
function button you reach the „Settings for measuring" menu (Fig. 48).
You use the [IEA/IPE] button to define whether or not substitute leakage
current measurement is to be used for determining the touch current.
Use the [RISO y/n] button to define whether the insulation resistance is to
be tested or not.
With the function button [hand/auto] you determine whether the test
procedure should be carried out manually (by hand) or automatically.
The headline (upper right) shows the setting:
Manually by hand (m), automatic (a)
DIN VDE 0751/EN 62353 at TG euro 1 med only
Fig. 46
The [cancel] button takes you back to the Main menu.
To move to the next stage press the [next] button (also see Diagram 1 in
Point 9.8).
Fig. 47
32
Manual test procedure means that you have to confirm every test
stage by pressing the [OK] button to get to the next test stage.
Automatic test procedure means that a test stage automatically
changes to the next after 5 seconds – except for confirmation of
connection to the mains voltage (see Point 6.6, basic settings of
the tester).
Fig. 48