GMW TG uni 1 Operating Manual page 37

Appliance tester for testing protective measures
Hide thumbs Also See for TG uni 1:
Table of Contents

Advertisement

In the „Memory menu" (Fig. 62) you can confirm or modify the following
parameters: device ID number, device designation, name of the inspector,
date and customer number.
You can also change the storage location – from internal memory to MMC/
SD card and vice-versa.
The current setting is displayed via a symbol in the headline of the memory
menu.
The appliance identification can also be read from a barcode label
using the optional barcode scanner. To do this, before starting up
the appliance tester connect the plug of the barcode scanner to the
RS232 socket of the tester and screw it tight.
You confirm the entries by pressing the [save] button – a „Data were
stored" message is displayed briefly and the display changes to Fig 63.
With the entry of an already used ID number, the data is „attached" as a
new test to this ID number. After saving, the display changes to the screen
for entering the ID number – the appliance tester is now ready for another
test (Fig. 63).
Fig. 61
Operating manual for the TG uni 1, TG euro 1 and TG euro 1 med appliance tester
Conducting tests in accordance with DIN VDE 0701-0702, Example 2 – Devices without a protective conductor (PC II)
Symbols for memory settings:
– Internal memory
– MMC/SD memory card
Fig. 62
The test results can be saved on a PC with the supplied PC program or
printed from there in the form of a test log. To do this, make the USB
connection between the appliance tester and the PC* (USB cable supplied
with the tester) or transfer the saved data from the memory card to the PC.
Procedure on the PC – see the operating instructions for the PC software.
* Switch the appliance tester off and on, leave the switch-on message as it is (Fig. 41).
Connect the USB cable. The appliance tester is automatically detected by Windows® as
removable media. The internal memory is assigned to the first free drive letter and the
memory card is assigned to the second free drive letter.
General information
Every test stage during which the test object is supplied with mains voltage
is time limited for reasons of safety.
The maximum duration of this test stage is approx. 5 minutes. This means
that the appliance tester will not remain in a state where the test object is
supplied with mains voltage for longer than approx. 5 minutes.
After this time, the mains voltage is switched off and an appropriate
message is shown in the display.
At this point you can continue the test with the [next] button or cancel it with
the [cancel] button.