DETERMINING PROPER SPAN GAIN (F2)
Note: Normally done from factory.
OVERVIEW
The Span Gain parameter found in F2 of the Setup Menu is directly related to the ADC (Analog to Digital Converter) integration time. This
means that the lower the setting, the higher the number of measurements per second. A span gain setting of 25 produces about 25 to 30
measurements per second, while a span gain of 200 produces only about 3 or 4 measurements per second.
There is really no wrong setting for span gain – except in two cases. Using a low setting for a high resolution, low output system could yield
instability. Using a high setting in a high output system could yield non-linearity.
SETTING THE INITIAL VALUE FOR SPAN GAIN
1. Determine the number of desired external graduations and choose the corresponding value listed in Table C-1 under the number closest to
your full-scale input range in millivolts.
2. Enter the Setup Menu and save this number for the Span Gain parameter in F2.
3. Perform a system calibration. If the calibration proves unsuccessful, or you wish to view the internal counts, proceed to the next set of
instructions.
VIEWING THE INTERNAL COUNTS
1. Enter the zero calibration menu (F16) and follow steps 1 to 3, but do not save the zero point.
2. After pressing ZERO to zero the offset, place the test weight(s) on the platform. The displayed count is the internal count. If the count
remains on zero, check your load cell connections.
3. At full scale, the displayed count should be a minimum of 2 times the desired external graduations. However, for maximum stability, a ratio
of 6:1 or higher is recommended.
4. If the displayed count is large enough, remove the test weight(s), re-zero the indicator if necessary, and proceed with the calibration. If the
displayed number is not large enough, increase the Span Gain to the next highest choice in the Setup Menu and re-calibrate.
External
Grads
0.2
500
1,000
150
1,500
200 100
2,000
2,500
3,000
4,000
5,000
6,000
8,000
10,000
12,000
15,000
20,000
30,000
40,000
0.4
0.6
0.8
75
50
25
25
75
50
50
75
50
– 150
75
75
– 200
100
75
– 200
150
100
150
150
200
150
200
Full Scale Input Range (mV/V)
1.0
1.2
1.4
1.6
25
25
25
25
25
25
25
25
50
50
25
25
50
50
50
50
75
50
50
50
75
75
50
50
100
75
75
75
150
100
100
75
150
150
100
100
200
150
150
150
200
200
150
200
200
1.8
2.0
2.2
2.4
25
25
25
25
25
25
25
25
25
25
25
25
25
25
25
25
50
50
50
25
50
50
50
50
50
50
50
50
75
75
75
50
75
75
75
75
100
100
100
75
150
150
150
100
150
150
150
150
200
200
200
150
200
Manual page 8
2.6
2.8
3.0
25
25
25
25
25
25
25
25
25
25
25
25
25
25
25
50
25
25
50
50
50
50
50
50
75
50
50
75
75
75
100
100
75
150
100
100
150
150
150
200
200
150