Vetek TI-1200 Complete Manual page 12

Weighing indicator
Table of Contents

Advertisement

KEY-IN ZERO CALIBRATION VALUE (F19)
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss. A valid zero calibration value,
obtained from a successful F16 calibration procedure, must be used.
1. While in the Setup mode, scroll to "F 19", then scroll down once using the ZERO key.
2. The display will momentarily show "CAL 0", followed by a flashing zero. Use the four directional keys to adjust the
displayed value to the zero calibration value.
3. After setting the exact value, press the NET/GROSS key to save the value.
4. The display will show "E CAL 0" momentarily, then revert back up to F19.
KEY-IN SPAN CALIBRATION VALUE (F20)
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss. A valid span calibration value,
obtained from a successful F17 calibration procedure, must be used.
1. While in the Setup mode, scroll to "F 20", then scroll down once using the ZERO key.
2. The display will momentarily show "CAL 1", followed by a flashing zero. Use the four directional keys to adjust the
displayed value to the span calibration value.
4. After setting the exact value, press the NET/GROSS key to save the value.
If the entered value is greater than zero, the display will show "E CAL 1" momentarily, then revert back up to F20. If a
5.
value of zero is entered, the indicator will briefly show "Err 5", then revert back to the screen described above in Step # 2.
Manual page 12