SETUP MENU DESCRIPTIONS AND PROCEDURES
Factory-set defaults are shown in bold
NAME/CODE
Specifies number of full-scale graduations. Value should be consistent with legal
F1
Graduations
requirements and environmental limits on the useful system resolution.
F2
Span Gain is related to A/D integration time. The larger the span gain, the higher the
Span Gain
internal resolution, but the slower the update speed. Note that the scale must be re-
calibrated whenever this parameter is altered. See Appendix C for more information.
F3
Selects the range within which the scale will automatically zero. Note that the scale
Zero Track Band
must be in standstill to automatically zero. Selections are in Display Divisions.
F4
Selects the range within which the scale may be zeroed. Note that the indicator must
Zero Range
be in standstill to zero the scale.
F5
Sets the level at which motion is detected by comparing the present display update
Motion Band
with the previous one. If motion is not detected for two seconds or more, scale is in
standstill and can process a Print or Zero command. Maximum value varies depending
on local regulations.
F6
Averages weight readings to produce higher stability. The higher the filter setting, the
Digital Filter
greater the stability but the slower the indicator's response time. Choose 8 unless a
very fast response is needed.
F7
Selects the desired formula which determines the point at which the indicator shows
Overload Limit
overload. All selections are based on the primary unit selected in F8. "FS" = Full scale
in primary units.
Selects the primary base unit to be used in the calibration process. Also the default unit
F8
Calib. Unit
for normal operation.
"1" = primary unit is lb.
F9
Determines the desired weight increments. Value should be consistent with legal
Display
requirements.
Divisions
F10
Determines location of the decimal point.
Decimal Pt.
F16
Places indicator into the zero calibration routine. Scrolling down with the ZERO key
Zero Calibration
one level begins the procedure.
F17
Places indicator into the span calibration routine. Scrolling down with the ZERO key
Span Calibration
one level begins the procedure.
Actuates the function that allows you to view both the zero and span calibration value.
F18
View Calibration
The values displayed in this function are valid only after Calibration (F16 & F17) has
been successfully completed. Scrolling down with the ZERO key one level begins the
procedure.
F19
Allows you to key-in known zero calibration value in case of memory loss in the field.
Key-in Zero
Scrolling down with the ZERO key one level begins the procedure.
F20
Allows you to key-in a known span calibration value in case of memory loss in the
Key-in Span
field. Scrolling down with the ZERO key one level begins the procedure.
F21
This sub-menu will reset all parameters in the "F" and "A" menu to the default
Factory Reset
settings. USE WITH CAUTION!
.
DESCRIPTION
"2" = primary unit is in kg.
CODE/VALUE
500
1,000
1,500 2,000
2,500
3,000
4,000 5,000
6,000
8,000
10,000 12,000
20,000 30,000 40,000 50,000
25
50
75
100
150
200
0d
0.5d
1d
3d
100% (US)
1.9% (CE)
2%
0.25d (CE)
1d (US) 3d
5d
10d
1
2
4
8
FS
FS + 2% (US)
FS + 9d (CE)
FS + 1d
1 (US)
2 (CE)
1
2
5
0
0.0
0.00
0.000
0.0000
00
Press ZERO key to begin
sequence
Press ZERO key to begin
sequence
Press ZERO key to begin
sequence
Press ZERO key to begin
sequence
Press ZERO key to begin
sequence
Press the ZERO key twice to
execute.
Manual page 7