Installing the IMA-x1000
4.6

Mounting the Power Supply Unit

Use M3 screws with the appropriate length (see Fig. 19,
p. 14) through the base mounting holes. This is neces-
sary to ensure a safety distance between the screw and
internal components.
Recommended mounting tightening torque is 0.6 Nm
(5.3 lb-in).
Chassis of the PSU
Mounting screw
L
L
min. 2 mm (0.078 in)
max.
4 mm (0.156 in)
Fig. 19: Mounting the Power Supply Unit
14
Mounting accessory
Washer spring
Washer