Appendix; Curves - Delta IMA-1000 Series Application Note

For medical and industrial applications
Table of Contents

Advertisement

Appendix

6.
Appendix
6.1

Curves

IMA-x1000
Output power [%]
100
80
80
90
Fig. 26: Output power versus input voltage
IMA-x1000-12
Output current [A]
84
69.4
9.6
Fig. 28: Output current versus output voltage 12 V
IMA-x1000-48
Output current (A)
21
17.8
38.4
Fig. 30: Output current versus output voltage 48 V
18
275
Input voltage [Vac]
12
14.4
Output voltage [Vdc]
48
56
Output voltage (Vdc)
IMA-x1000
Tested at 90 Vac input
Output power [%]
100
50
-20
45
50
-4
113
122
standard mounting orientation
all other mounting orientations
Fig. 27: Output power versus ambient temperature
IMA-x1000-24
Output current [A]
42
34.7
19.2
24
Fig. 29: Output current versus output voltage 24 V
IMA-x1000-xx-Series_Application_Note_V4.0_XX_EN_2019-08-19
65
70
[°C]
149
158
[°F]
Ambient temperature
28.8
Output voltage [Vdc]

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ima-1000-12 seriesIma-1000-24 seriesIma-1000-48 series

Table of Contents