Download Print this page

Advertisement

Quick Links

MODIFIED CALIBRATION GUIDE FOR SI-9001/9002 (Offset):
Explanation of adjustment point. Please refer to Fig. 7.
(1) SW1: Switch for selecting attenuation ratio.
(2) SW2: Power switch.
(3) VR4: Variable resistor for adjusting offset.
(4) VR2: Variable resistor for adjusting offset.
(5) VR1: Variable resistor for adjusting low frequency CMRR.
(6) VR3: Variable resistor for adjusting accuracy.
(7) VC1: Variable capacitor for adjusting square wave compensation.
(8) VC2: Variable capacitor for adjusting square wave compensation.
(9) CON1: Power jack for connecting to 6VDC power adapter
Adjustment Procedure:
Please connect a 6VDC power adaptor to CON1 as shown in the Figure above, and
turn "on" SW2, the LED indicator must light up. Keep the unit turned "on" for 10
minutes to keep it warm before the calibration starts. Then follow the steps below.
Offset Adjustment:
A. Referring to the Figure below, connect both input lines (black and red) together,
thus actually shorting the two input-leads. Make sure they are shorted to ensure that the
unit is calibrated properly. (Use an alligator clip for contact on the input lead, if
possible).
B. Connect BNC connector output to a DMM as shown in the Figure below. The
DMM should be properly set to mV range, better if the DMM have an auto-range
capability.
C. Put SW1 at 1/10 (or 1/20) attenuation setting as indicated at the front plastic
cover of the probe.
D. By referring to the Figure above, adjust VR2 using a ceramic screw or any tools
appropriate for adjusting variable component to make the output voltage approach to
zero. (This way the Offset is adjusted into range within 0.1mV).
E. Put SW1 at 1/100 (or 1/200) attenuation setting as indicated again at the front

Advertisement

loading

Summary of Contents for Testec SI-9001

  • Page 1 MODIFIED CALIBRATION GUIDE FOR SI-9001/9002 (Offset): Explanation of adjustment point. Please refer to Fig. 7. (1) SW1: Switch for selecting attenuation ratio. (2) SW2: Power switch. (3) VR4: Variable resistor for adjusting offset. (4) VR2: Variable resistor for adjusting offset.
  • Page 2 plastic cover of the probe. F. Again referring to the Figure above, adjust VR4 to make output voltage approach to zero. (This way the Offset is adjusted into range within 0.1mV). However, when VR4 is changed, the Offset at 1/10 is affected, thus we must calibrate again the 1/10 by repeating steps C and D.
  • Page 3 Calibration of TT-SI 9001 & 9002 1. How to open the case, please follow the steps. 1.1 Remove the battery lid. 1.2 You will see the two screws and loosen them. Fig. 1 1.3 Near the BNC cable side, there are two protuberant mortises. Use fingers to push back and pull upward.
  • Page 4 1.5 Hold the bottom case and push back (to input lines direction). Fig.4 1.6 You will see the two metal cases. There is one screw on the PCB. Loosen it, then the whole metal case can be taken out. There are two wires still connected with the upper case. Leaves them there. Fig.
  • Page 5 CALIBRATION GUIDE FOR SI-9001 / SI-9002 (Offset): Explanation of adjustment point. Please refer to Fig. 7. (1) SW1: Switch for selecting attenuation ratio. (2) SW2: Power switch. (3) VR4: Variable resistor for adjusting offset. (4) VR2: Variable resistor for adjusting offset.
  • Page 6 D. By referring to the Figure above, adjust VR2 using a ceramic screw or any tools appropriate for adjusting variable component to make the output voltage approach to zero. (This way the Offset is adjusted into range within 0.1mV). E. Put SW1 at 1/100 (or 1/200) attenuation setting as indicated again at the front plastic cover of the probe.
  • Page 7 1.8 Low Frequency CMRR Adjustment a. Referring to Fig. 9, connect output BNC to the oscilloscope which has good grounding. b. Connect both input lines to a 220/200VAC, 60/50Hz power source. c. Put SW1 at 1/10 (or 1/20). d. Set the VOLTS/DIV switch on oscilloscope’s panel to 1mV and the TIME/DIV switch to 5ms. Then a 60/50Hz sinusoid waveform should be displayed on the screen.
  • Page 8 1.9 Square Wave Adjustment a. Referring to Fig. 10, connect output BNC to the oscilloscope has good grounding. b. Connect the red input line to a 20Vp-p, 10KHz square wave signal source and the black input line to ground. c. Put SW1 at 1/10 (or 1/20). d.
  • Page 9 Fig. 11 1.10 Repeat 3.2 and 3.3 adjustment until both adjustments are O.K. 1.11 High Frequency CMRR adjustment a. Referring to Fig. 12, connect output BNC to the oscilloscope has good grounding. b. Connect both input lines to a 155Vp-p, 20KHz, sine wave signal source. c.
  • Page 10 Fig. 12 1.12 Accuracy Adjustment a. Referring to Fig. 13, connect output BNC to a DMM. b. Connect the red input line to the positive terminal of a DC voltage source (70VDC) and the black input line to the negative terminal source (70VDC). c.
  • Page 11 d. Adjust VR3 to make the output voltage equal to 6.9V (or 3.51V for 1/20). Fig. 13...

This manual is also suitable for:

Si-9002Tt-si 9001Tt-si 9002