The TRIGger:COUNt? <slx> command queries the specified scanning list timer cycle count.
<slx>
SL1, SL2, SL3, SL4
Returns the cycle count value (1 to 9999), or "CONT", if the value is continuous.
TRIGger:DELay <hh:mm:ss>
TRIGger:DELay <slx>, <hh:mm:ss>
The TRIGger:DELay <hh:mm:ss> command is used to set the general timer cycle delay time.
The TRIGger:DELay <slx>, <hh:mm:ss> command is used to set the individual scanning list timer cycle delay time.
<slx>
SL1, SL2, SL3, SL4
<hh:mm:ss> 00:00:00 to 99:59:59
Terminates the current measurement cycle and scanning routine, and sets the trigger mode to HOLD.
TRIGger:DELay?
TRIGger:DELay? <slx>
The TRIGger:DELay? Command queries the general timer cycle delay time.
The TRIGger:DELay? <slx> command queries the specified scanning list timer cycle delay time.
<slx>
SL1, SL2, SL3, SL4
Returns the delay time (00:00:00 to 99:59:59).
SAMPle Group
SAMPle:COUNt?
SAMPle:COUNt? <slx>
The SAMPle:COUNt? command queries the general timer sample count. This is the same as the general timer cycle
count.
The SAMPle:COUNt? <slx> command queries the specified scanning list sample count. The sample count is the
product of the timer cycle count and the number of active channels in the specified scanning list.
<slx>
SL1, SL2, SL3, SL4
Returns the sample count (1 to 9999), or "CONT", if the value is continuous.
SAMPle:DELay <slx>, <hh:mm:ss>
This command is used to set the specified scanning list timer sample delay.
<slx>
SL1, SL2, SL3, SL4
<hh:mm:ss> 00:00:00 to 99:59:59
DEFault
Changing a timer sample delay for a scanning list that is currently in use will result in the instrument terminating the
current measurement cycle and scanning routine. Also, the Run/Hold trigger mode will be set to HOLD.
Specify scan list
Specify scan list
Specify timer delay
Specify scan list
Specify scan list
Specify scan list
Specify trigger delay
Trigger delay 00:00:00
Page 29 of 44
TTI-7 Iss.02 – 06/11