Installation

NOTICE
Do not over tighten TLM into tank bung!
Over tightening can cause permanent
damage and result in inaccurate readings.
Do not use thread sealant or adhesive!
Many of these products are chemically
incompatible with the ABS plastic. PTFE
tape is acceptable.
Do not locate TLM next to the tank wall.
The wall can interfere with the sensor's 30
degree cone angle.
Do not install the TLM in the tank's fill
port. Repeated removal of the TLM will
damage the unit and void the warranty.
1.
Remove the bung fitting and screw in the
TLM hand-tight.
Do not use a wrench to tighten as
this may damage the TLM.
Do not use thread lock adhesive as
this may damage the TLM.
2.
Mount the TLM as close as possible to
the center of the tank.
3.
Do not mount any fittings or T connects
between TLM and the tank.
3A5411J
NOTICE
The TLM will not operate correctly if tilted
more than 2° from the surface of the tank
liquid.
The TLM will not operate correctly if the
TLM is mounted less than 5 in. (127 mm)
from the maximum possible fill height of the
fluid being measured.
The TLM will not read properly if:
Fluid in the tank is agitated (e.g. filling
a tank with oil or anti-freeze). Be sure
to take all TLM readings when the fluid
in the tank is calm.
The TLM is mounted next to the tank
wall or in a corner.
If there is any structural bracing or
obstruction between the TLM and the
surface of fluid in tank.
TLM must be mounted within 2° from per-
pendicular to the surface of the fluid.
F
. 5
IG
Installation
7