Installation and Configuration in S Mode
Table 3-2. Configuration Parameters for Switches 6 through 8
!
6
!
!
OFF
!
OFF
!
OFF
!
OFF
!
ON
!
ON
!
ON
!
ON
!

Default Settings for Configuration Switch

Figure 3-2 shows the GPIB-232CT's factory default setting for switch U22.
Switch 1 is OFF; this indicates that the GPIB-232CT will be operating in S
mode. Switch 2 is OFF; this indicates that the GPIB-232CT will be using 7
bits/character for its serial data transfers. Switch 3 is OFF; this indicates 1
stop bit per character. Switches 4 and 5 are both OFF; this indicates that
parity is disabled. Switches 6 through 8 are in the ON OFF ON positions,
respectively, indicating that the serial port will be operating at 9600 baud.

Figure 3-2. Factory Default Settings for Switch U22