COMMON OPTIONS AND ACCESSORIES
Hospital Grade Receptacle
WARNING:
OUTLET IS FOR MEDICAL EQUIPMENT ONLY.
CAUTION: CHECK GROUND CONTINUITY PERIODICALLY.
120V MAXIMUM OUTPUT - 5.0A, 120V
220V MAXIMUM OUTPUT - 2.5A, 220V
WARNING:
DO NOT DISCARD COVER. COVER IS TO BE REAPPLIED AFTER REMOVAL OF EQUIPMENT FROM
MULTIPLE SOCKET OUTLET.
WARNING:
MULTIPLE SOCKET OUTLET LOCATED ON SIDE OF TABLE IS TO BE USED FOR MEDICAL EQUIPMENT ONLY
(WHERE SAFETY CERTIFICATION HAS BEEN PERFORMED IN ACCORDANCE TO IEC 60601-1 AND/OR IEC
60601-1-1). USE OF EQUIPMENT NOT COMPLYING WITH THE EQUIVALENT SAFETY REQUIREMENT OF THIS
EQUIPMENT MAY LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING SYSTEM. THE MULTIPLE SOCKET
OUTLET LOCATED ON THE BACK SIDE OF THE PANEL, ON THE INSIDED OF THE TABLE, IS TO BE USED FOR
THE POWER SUPPLY FOR THE COLUMN ONLY, THIS OUTLET SHOULD NOT BE USED FOR ANY OTHER ITEMS.
WARNING:
GROUND RELIABILITY CAN ONLY BE ACHIEVED WHEN POWER SUPPLY IS CONNECTED TO AN EQUIVALENT
RECEPTACLE MARKED "HOSPITAL ONLY" OR "HOSPITAL GRADE". WARNING: ADDITIONAL EXTENSION CORD OR
MULTI-SOCKET OUTLET IS NOT TO BE CONNECTED TO TABLE.
Service Note: If the maximum load of the receptacle is exceeded, two circuit breakers will interrupt power. To reset, toggle the far two
circuit breakers located behind the receptacle outlet box in the back of the drawer panel.