UMF Medical 5005 Service Manual

UMF Medical 5005 Service Manual

Bariatric power treatment table

Advertisement

Quick Links

5005 Bariatric Power Treatment
Table Service Manual
UMF Medical
1316 Eisenhower Blvd.
Johnstown, PA 15904
(814) 266-8726
12/2011
WWW.UMFMEDICAL.COM

Advertisement

Table of Contents
loading

Summary of Contents for UMF Medical 5005

  • Page 1 5005 Bariatric Power Treatment Table Service Manual UMF Medical 1316 Eisenhower Blvd. Johnstown, PA 15904 (814) 266-8726 12/2011 WWW.UMFMEDICAL.COM...
  • Page 2: Table Of Contents

    3.4 ONE OR MORE OF THE ACTUATORS MAKES A GRINDING OR OTHERWISE LOUD NOISE WHEN ACTIVATED. WHAT SHOULD I DO? ..7 4 REPLACEMENT PARTS LIST AND EXPLODED VIEW ..............................9 5 REPLACEMENT PART ORDERING INFORMATION ..............................10 6 EQUIPMENT CARE ........................................12 UMF Medical www.umfmedical.com...
  • Page 3: Important Information

    1 IMPORTANT INFORMATION 1.1 READ THIS FIRST! 1.3 CLASSIFICATIONS 1.1.1 Safety First: The primary concern of UMF Medical is Protection against electric shock: Class I, Class B Applied Parts that this equipment is operated and maintained with Protection against harmful ingress of water: Ordinary the safety of the patient and doctor in mind.
  • Page 4: Warranty Information

    United Metal Fabricators Inc. d.b.a. other warranty is applicable to this product. The duration of any UMF Medical shall not be liable for changes in color caused by implied warranties, including the implied warranty of natural or artificial light, matching of color, grain or texture merchantability, is limited to the duration of the express except to commercially acceptable standards.
  • Page 5: Troubleshooting And Repair

    3.1.5 Check the outside of the control box for excessive Remove heat. If overheated, unplug power cord and allow the 3.1.7.3 Remove the four (4) screws on the installation unit to cool for approx. 30 minutes. bracket and remove the Control Box. UMF Medical www.umfmedical.com...
  • Page 6: The Powered Back Section Is Not Moving When Activated. What Should I Do

    3.2.3 Check plugs. Check all plugs and connections to ensure Pins that all wires are properly connected. UMF Medical www.umfmedical.com...
  • Page 7: The Table Is Not Moving Up And Down. What Should I Do

    3.2.6 Replace hand control. See section 3.1.8. could cause resistance to motion. 3.2.7 If this does not work, contact UMF Medical Customer 3.4.2 Replace the appropriate actuator. Service at (814)266-8726.
  • Page 8 The table is very heavy and must to the Control Box and secure in place. be braced properly before removing the 3.4.3 If this does not work, Contact UMF Medical actuator to ensure that the table does not fall. Customer Service at (814)266-8726.
  • Page 9: Replacement Parts List And Exploded View

    4 REPLACEMENT PARTS LIST AND EXPLODED VIEW Accessory Rails (2x) Upholstered top (includes the foot section) Paper roll Actuator and holder attachment bracket Top Cover Footstep assembly Frame Drawer assembly Actuator Control Box 2 function Hand Control Actuator Base Base UMF Medical www.umfmedical.com...
  • Page 10 #375 Safety rails (pair) requires 2 each #039 accessory rails. NPN Actuators #800 Paper cutter band NPN Control box #801 T- cable #063 2-function Hand Control #810 Restraint straps (set of 2) requires #039 accessory rails. #069 2-function Foot Control UMF Medical www.umfmedical.com...
  • Page 11: Replacement Part Ordering Information

    5 REPLACEMENT PART ORDERING INFORMATION To order replacement parts, please obtain the serial number of the table and contact your authorized UMF Medical distributor. If you are unsure of your authorized distributor, please contact UMF Medical customer service at: UMF Medical...
  • Page 12: Equipment Care

    Care should be used where solvents or dyes are used. UMF Medical recommends the use of Murphy’s oil soap or Ivory liquid. If dirt or stains still remain, use rubbing alcohol.