Table of Contents

Advertisement

MANUAL - SOLVENT RECYCLERS - URS1600 SERIES, URS2000 SERIES & RI 80 Rev 2019-05
SETUP
To change settings, enter Setup Mode: while pressing SETUP, press START. "SET-UP MODE" is dis-
played. To change the temperature units from °C to °F, press and hold the + and - keys together
for 2 seconds.

STEP SETTING

1
SET-PT = XXX°C (°F)
2
POWER = XXX%
3
SHUT-OFF = AUTO**
4
BAKE TIME = XXM
**WARNING: SHUT-OFF TIME MUST NOT BE SET TO MORE THAN 12 HOURS FOR THE URS2000
ADJUST ACCEPT DESCRIPTION
To choose a SET POINT, look up the solvent's BP
+ OR -
OK
(boiling point (MSDS, online etc) and add about
40°C (100°F).
If boiling is too vigorous due to one or more of the
following conditions, decrease POWER by one
increment or more.
• vapour leaks at the Lid Gasket
+ OR -
OK
• recycled solvent comes out too hot
• waste material is carried into the recycled sol-
vent
Otherwise, use 100%.
If too much solvent remains in the Distillation Tank
after recycling and the problems in the Trou-
blshooting Guide have been ruled out, over-ride
+ OR -
OK
the AUTO SHUT-OFF and manually select a heat-
ing time (eg: 4 hours). The heater will stay on for
this amount of time and then cooling will begin.
Only available when SHUT-OFF = AUTO. In-
+ OR -
OK
crease if residue (puck) is too wet.
SERIES AND THE RI 80
8

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents