Chapter 5
*ESE
Event Status Enable
Command
Syntax *ESE mask_value
Example *ESE 1
Returns None
Programming
Example
© 2010 Colby Instruments. PDL-100A Operating and Programming Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The *ESE command sets the Standard Event Status
Enable Register bits. The ESR Register contains a
mask value for the bits to be enabled in the Standard
Event Status Register. A one in the ESR mask will
enable the corresponding bit in the ESR to be
reported, a zero will disable the bit. For example, a
mask value of 255 will enable all bits while a mask
value of 1 will enable only the OPC bit to be set.
Where mask_value is
Bit #
Value Enables
PON – Power on (not used)
7
128
URQ – User Request (not used)
6
64
CME – Command Error
5
32
EXE – Execution Error
4
16
DDE – Device Dependent Error
3
8
QYE – Query Error
2
4
RQC – Request Control (not used)
1
2
OPC – Operation Complete
0
1
HP : OUTPUT 705;"*ESE 255"
NI : CWGPIB.Write "*ESE 1"